Portable Meter Kit, 24 L x 10 W x 6.5 cm H, Conductivity/TDS/Salinity/Resistivity with Temperature Measuring Modes, -5 to 105°C/22 to 221°F Temperature Range, 0.001 to 3000 mS/cm Conductivity Range, 0.001 to 200 ppm TDS Range
Get high accuracy and premium performance everywhere you need it with the Thermo Scientific™ Orion Star™ A322 conductivity portable meter, designed for a wide range of conductivity, total dissolved solids (TDS), salinity, resistivity and temperature testing and field applications. Evaluate samples in the most demanding locations using this waterproof meter with IP67-rated housing. Select linear, non-linear pure water or EP temperature compensation with 5, 10, 15, 20 or 25°C reference temperatures. Log up to 5000 data point sets with time/date stamp and transfer to a printer or computer. Take your measurements at the source with our premium line of portable meters. The Thermo Scientific™ Orion Star™ A322 conductivity portable meter offers an information-rich screen and user-friendly operation. This rugged, waterproof portable meter is ready to tackle the elements with you. Orion Star™ A322 conductivity portable meters are designed to be easy to use, dependable field meters that offer advanced features and functions including a 5000 point date log with sample ID and user ID options, calibration and high/low limit alarms and password-protected measurement methods.
Dimension | : | 24 L x 10 W x 6.5 cm H |
Alarm Output | : | High/Low Limit Alarm, Calibration Due Alarm |
Backlight Option | : | Yes, Selectable |
Calibration | : | Conductivity with Calibration Editing Option and Temperature |
Cell Constant | : | 0.001 to 199.9 cm-1 |
Compatible Probes | : | 2-cell or 4-cell Conductivity Sensors with Built-in Temperature |
Conductivity Accuracy | : | 0.5% of Reading ±1 Digit > 3 µS, 0.5% of Reading ±0.01 µS = 3 µS |
Conductivity Calibration Points | : | Up to 5 |
Conductivity Range | : | 0.001 to 3000 mS/cm |
Conductivity Resolution | : | 0.001 µS Minimum, Auto Ranging upto 4 Significant Digits |
Connector Type | : | Input: 8-pin MiniDIN (Conductivity Sensor with Built-in Temperature), Output: USB, RS232 |
Data Management Details | : | Transfer Single, Range or All Data Points to Printer or Computer |
Display Type | : | Graphic LCD with Backlight |
Includes | : | Starâ„¢ A322 Conductivity Meter, 013016MD Orion Pure Water (K=0.1) Conductivity Sensor with Flow-cell, 011008 Orion 100 µS Conductivity Standard, 911110 Orion Rinse Solution Pouch 10 Pack, Protective Meter Armor, Hard-sided Field Case, 4 AA Batteries, Literature CD, Printed Quick Start Guide, Computer Interface Cable and Meter Test Certificate |
Keypad | : | Comprehensive with Menu-specific Function Keys and Dual Purpose Scroll/Shortcut Keys |
Languages | : | English, Spanish, French, Italian, German and Chinese, Portuguese and Korean Available with Software Update |
Measuring Modes | : | Conductivity/TDS/Salinity/Resistivity with Temperature |
Method | : | 10 per Channel with Password Protection |
Number of Channel | : | 1 |
Operating Temperature | : | 5 to 45°C (ambient) |
Power Supply | : | 4 AA Batteries 800 Hours of Life or Universal AC Adapter 50 to 60 Hz, 100 to 240 VAC Power |
Probe Type | : | Conductivity sensor (Includes Orion Pure Water 2-cell) |
Relative Humidity Range | : | 5 to 85%, Non-condensing |
Resistivity Accuracy | : | 0.5% of Reading ±1 Digit |
Resistivity Range | : | 2 ohm to 100 meg-ohm |
Resistivity Resolution | : | 1 ohm or 0.1 meg-ohm, Auto Ranging |
Salinity Accuracy | : | 0.5% of Reading ±1 Digit |
Salinity Measurement Modes | : | Practical Salinity (psu) or Natural Sea Water (ppt) |
Salinity Range | : | 0.06 to 80 psu, 0.05 to 42 ppt |
Salinity Resolution | : | Minimum 0.01 psu or 0.01 ppt, Auto Ranging |
Stability | : | Selectable Fast, Medium, Slow or Smart Stability Setting, Selectable off or Automatic Smart Averaging Setting |
TDS Accuracy | : | 0.5% of Reading ±1 Digit |
TDS Factor Range | : | 0.02 to 9.99, Linear |
TDS Range | : | 0.001 to 200 ppm |
TDS Resolution | : | Minimum 0.001 ppm, Auto Ranging up to 4 Significant Digits |
Temperature Accuracy | : | ±0.1°C |
Temperature Compensation | : | Linear, nLFn (Non-linear Ultra Pure Water), nLFu (Non-linear Ultra Pure Degassed Water), EP (Off with Warning) or Off |
Temperature Probe Calibration | : | 1 Point Temperature Offset |
Temperature Range | : | −5 to 105°C, 22 to 221°F |
Temperature Resolution | : | 0.1°C, 0.1°F |
Temperature Selection | : | Manual or Automatic with ATC Temperature Probe |